4.12 Configuration screen

The password is made up of 6-digit number and *.
Note: If the password is set wrong, the user will not be able to enter the configuration screen.

The factory default password is 000000.

The following will introduce each configuration screen:

4.12.1 System configuration

Language selection: Simplified Chinese and English are optional.

User password: Users can set passwords according to their needs.

Record Name: Users can set record’s name according to their own needs, factory default:

paperless recorder.

Date &time: display "Year- Month – Day Hour – Minute- Second", it can modify and

adjust time.

Cold end adjustment: adjust and display the temperature at cold end.

Treatment of interruption: low limit of measurement range, high limit of measurement

range, maintain the preceding value, - - optional.

Correspondence address: Address range: 1~255.

The baud rate : 1200, 2400, 4800, 9600, 19200, 38400, 57600 are optional.

IP address: The IP address of

Modbus Tcp

.

Port: The port number of

Modbus Tcp

.

Summary of Contents for SUP-5000C

Page 1: ...Color Paperless Recorder Specification Paperless recorder SUP 5000C...

Page 2: ...6 4 3 Real time curve screen 6 4 4 Bar chart screen 8 4 5 Digital display screen 9 4 6 Historical curve screen 10 4 7 Alarming list screen 11 4 8 File list screen 12 4 9 Printing screen In case there...

Page 3: ...A 4 20mA 0 10mA radication 4 20mA radication Voltage 0 5V 1 5V 0 5V radication 1 5V radication 0 100mV 20mV 100mV Thermal resistance Pt100 Cu50 Cu53 Cu100 BA1 BA2 Linear resistance 0 400 Thermal resis...

Page 4: ...ction Memory length UPS Continuous recording 24 days while the interval is 1 second 5825 days while the interval is 240 seconds Formula Recording duration day 64 1024 1024 Recording interval s No of c...

Page 5: ...Recommend the use of our products When using storage media at high temperature about above 40 please insert the storage media when saving the data After the data is saved remove it and keep it Do not...

Page 6: ...iances it will have a bad impact on the interior of the instrument Places where there are more oil smoke steam moisture dust corrosive gas and so on Oil fume steam moisture dust corrosive gas and so o...

Page 7: ...unting brackets on both sides of the instrument 2 Installation figure is shown as the right figure Wiring diagram for instrument Terminal diagram of instrument tail Wiring illustration Functions and o...

Page 8: ...h running screen File record mark Display available It indicates the file is circularly recording Display available with arrow mark It indicates the file is not circular recording Display not availabl...

Page 9: ...ient for users to estimate time and data value The time represented by the current grid Curve combination Display the name of the current curve combination each curve combination can include six curve...

Page 10: ...ing to the display channel can be set As for the setup method please refer to Section 4 12 4 1 Input method of channel s bit number Bar chart The length of the bar chart s ruler is 10 grids and the fi...

Page 11: ...l Unit It displays the data unit of the channel and set it Please refer to Section 4 12 4 2 Input Method of Channel s Unit Over limit alarming instructions From top to bottom each block in turn is the...

Page 12: ...3 Real time Curve Screen Data curve Maximum 6 curves can be displayed on the same screen at the same time There are 6 kinds of curve colors which can be set by display configuration Grid convenient f...

Page 13: ...up Press the Time to modify the recalling time Year Month Day Hour Minute Second 4 7 Alarming list screen It displays the alarming information of channels and output status of the relay with maximum 1...

Page 14: ...after the limit is up to the new alarming records will cover the earliest alarming records Serial number The serial number of the alarming is generated and arranged in a time manner The closer the oc...

Page 15: ...mp to the backup screen you can back up a single historical file or all historical files For the specific operations you can view the 4 10 Backup Screen and press the Backspace key to back to the file...

Page 16: ...n press the Enter key to modify the cursor and then press the Enter key to confirm exit after modification Press the Print key to start printing Press Delete key to delete printing Connection Schemati...

Page 17: ...a 4 10 Backup screen In case there is the backup function the screen is available The paperless recorder used flash or SD card to back up and transfer the recorded data in the instrument The USB inter...

Page 18: ...tarting and end time of the curve it will appear the error prompt File name Select the file name of the backup data in the flash When the backup content chooses all files the default prefix of folder...

Page 19: ...nd total power off duration which can save maximum 100 alarming messages When the number of saved items is up the new power off record files will cover the earliest records Serial number The Serial nu...

Page 20: ...s according to their needs Record Name Users can set record s name according to their own needs factory default paperless recorder Date time display Year Month Day Hour Minute Second it can modify and...

Page 21: ...e box Do you save the modified parameters In the dialog box select Yes to save the modified parameters and No to save the modified parameters 4 12 2 Record configuration Record mode non circulation mo...

Page 22: ...Every time after the power on is started the instrument automatically creates new files and records data Alarming trigger It can be set to start data storage record when a certain alarming corresponds...

Page 23: ...l time curve screen Start using Select Yes the corresponding combination of curves can be displayed on the real time curve screen but select No there is no display Curve 1 to 6 Select the name of the...

Page 24: ...mV 0 20mA 0 10 mA 4 20mA 0 5V 1 5V 0 10mA 4 20mA 0 5V 1 5V nil The permitted ranges of all signals are stated as follows Signal model Scope of range Signal model Scope of range Signal model Scope of r...

Page 25: ...g the low limit value of range and set 0 to switch off the function of small signal cutoff Range 0 99 9 Linear correction The quantity of work is linearly corrected according to formula y Kx b where K...

Page 26: ...yboard operations It is easy to operate learn and use which it uses international coding to solve the input problem of the bit numbers and special units of Chinese character When the cursor moves to t...

Page 27: ...Select key to select display d in the display field of the phonetic letters combination 4 Press the keys Left shift Right shift and Up shift to move the cursor to select the phonetic letter a 5 Press...

Page 28: ...in the input field 6 Press the Complete key to exit the input method and complete the input of the bit number III Lowcase letter input method screen Input the lowcase letter a 1 Press the Enter key to...

Page 29: ...to select the 7 Press the Select key to select display 0 in the input field 8 Press the keys Left shift Right shift and Up shift to move the cursor to select the 1 9 Press the Select key to select di...

Page 30: ...ase letters while inputting lowcase letters Display16 special symbols while inputting special symbols Input method selection field where select the required input method here Operation keys Press the...

Page 31: ...o the software disc field with the special symbol input 5 Press the Select key to select display in the input field 6 Press the Complete key to exit the input method and complete the input of unit II...

Page 32: ...field 6 Press the Complete key to exit the input method and complete the input of unit IV Special units input method screen Input the special unit 1 Press the Enter key to enter the input method scree...

Page 33: ...peration In this screen Move the cursor by pressing the keys Left shift and Right shift Select the wanted contents by pressing the keys Down shift Up shift Page down and Page up Press the Enter key to...

Page 34: ...rocessing This instrument provides users with two standard interfaces RS 485 and EtherNet for communication with the upper computer RS 485 mode allows a upper computer to connect multiple recorders si...

Page 35: ...acts Long Read only No 0 11 bits represent the output states for No 1 12 contacts 0 disconnect 1 close Table2 0x03 Maintaining register address table that read command corresponds to 0x10 write comman...

Page 36: ...r address from low to high is 192 and 168 respectively 8 8 No 3 4 bytes of IP address Short Read write Same as last example 2 bytes of the register address from low to high is 100 and 2 respectively 9...

Page 37: ...ent Char Read write The value is 10 times after the magnification of the parameter 28 30 Signal cutoff Short Read write The value is 10 times after the magnification of the parameter 29 31 Linear corr...

Page 38: ...the parameters 5 minutes never 44 54 Circulate the display time Char Read write 0 3 respectively corresponds to the parameters 4 30 seconds 45 55 Display boot screen Char Read write 46 56 Serial numbe...

Page 39: ...corresponds to channel 1 12 Note 1 59 69 Alarming type for alarming trigger Char Read write 0 3 respectively corresponds to low low limit high high limit 60 70 Bit number of relay triggered by relay C...

Page 40: ...w limit alarming high limit alarming and high high limit alarming The four groups of alarming types share the same register address The alarming types register address 35 are used to determine the par...

Page 41: ...SD card 1piece When users place orders on SD card storage function SD card will be provided whose capacity depends on the order by users 35 35...

Reviews: