15
Time
Data
---------------------------------------------------
18-07-24 15:31:40 850 - -- measurement value of starting time
18-07-24 15:31:41 850
18-07-24 15:31:42 850
...
18-07-24 15:31:58 425
18-07-24 15:31:59 425
18-07-24 15:32:00 425 - -- measurement value of end time
---------------------------------------------------
B. Timing printing
The time interval and starting time for the timing printing are set in the system configuration.
When the system time reaches the set interval time, the instrument will automatically control the
printer for timing printing.
Timing printing format:
--------------------------
Printing time: 18-07-24 14:30:02 ------------------- date and time
Channel 01:100.0
℃
------------------- measurement value of No. 01 channel
Channel 02:100.0
℃
------------------- measurement value of No.02 channel
.....
.....
Channel 11:100.0
℃
------------------- measurement value of No.11 channel
Channel 12:100.0
℃
------------------- measurement value of No. 12 channel
--------------------------
Note: The manual printing can print data and curves, while the timing printing can only print
data.
4.10 Backup screen (In case there is the backup function, the screen is available)
The paperless recorder used flash or SD card to back up and transfer the recorded data in the
instrument. The USB interface is on the front panel of the instrument, and the SD card interface is
on the left side of the instrument. As long as the operation cover is opened, the flash or SD card
can be inserted into for the operation of data backup.
SD card has automatic backup function: After inserting SD card into the driver, the
instrument will automatically back up the recorded historical data to SD card at 0 o'clock every
day; When the recording is stopped or the triggering condition set by the user is reached, the
instrument will automatically back up the historical data to SD card.
Automatically back up file’s storage directory:
1. File’s storage directory backed up at 0:00 a day: TIMEAUTO (folder’s name)/instrument
date (year/month/day)/instrument time (hour/minute).
2. Record the file’s storage directory for stopping backup: RECAUTO (folder’s