8.0 Appendix
8.1 Default Settings For 1020i PLC
Input the values in Table 8-1 below for default optimum flow and position values.
Calibration should be performed after input, by adjusting OPVal when flowing calibrated plates.
Table 8-1. Default Optimum Flow and Position Values
Range
OPVal
Position
1
25
5266272
2
50
4874796
3
100
4350105
4
150
3945465
5
200
3595657
6
300
3002954
7
400
2504226
8
500
2065429
9
700
1299986
10
1000
342827
Table 8-2. Default PID Settings
Input
Value
Flow P
0.101
Flow I
0.580
Flow D
0.0
Pressure P
0.50
Pressure I
1.0
Pressure D
0.0
Table 8-3. Default Calibration Values (Select Enable All Calibration Values)
Input
Adder
Gain
Average
Input 1 Baro
11.5160
0.8702
30
Input 2 Test Pressure
0
20
30
Input 3 Flow Pressure
0
10
30
Input 4 Not Available
0
1
30