12
Register
Function
Data Type
Read/
Write
Value
Units
Scale
40019
PercentFlow
Int-16
R
Returns SCRT(FlowPress /
FullScaleFlowPress)
%
x10
40020
Swirl
Int-16
R
Not Used
40021
Frequency
Int-16
R
Not Used
40022
FullScaleFlow
Int-16
R
Returns the full-scale reading for the current
flow range.
CFM
x10
40023
RangeSetting
Int-16
R/W
Display current and set flow range 1-10
40024
TestPressureSetting
Int-16
R
Returns the current Test Pressure Set point
InH2o
x100
40025
FlowRateSetting
Int-16
R
Returns the current Flow Rate Set point
CFM
x10
40026
Leakage
Int-16
R/W
Sets leakage offset in CFM
CFM
x10
40027
FilteredData
Int-16
R
Raw AI1 Value
counts
0-65535
40028
FilteredData
Int-16
R
Raw AI2 Value
counts
0-65535
40029
FilteredData
Int-16
R
Raw AI3 Value
counts
0-65535
40030
FilteredData
Int-16
R
Raw AI4 Value
counts
0-65535
40031
FilteredData
Int-16
R
Raw Temp 1
counts
0-65535
40032
FilteredData
Int-16
R
Raw Temp 2
counts
0-65535
40033
FilteredData
Int-16
R
Not Used
counts
0-65535
40034
FilteredData
Int-16
R
Not Used
counts
0-65535
40035-36
TestPressureSetFloat
Float-32
R/W
Floating point value for test pressure setpoint InH2o
None
40037-38
FlowRateSetFloat
Float-32
R/W
Floating point value for flow rate setpoint
CFM
None
40039-40
ReadyToTest
Float-32
R
0=NotReady 1=Ready/Stable
40041-42
SerialNumber
Float-33
R
SerialNumber of flow bench
40043-44
MotorSpeedCmd
Float-32
R
Returns current VFD speed command in %
%
None
40045-46
VFDFault
Float-32
R
0=Normal Operation 1=VFD in fault
40047-48
TestPressureFloat
Float-32
R
Floating point value for test pressure
InH2o
None
40049-50
FlowPressureFloat
Float-32
R
Floating point value for differential pressure
InH2o
None
40051-52
FlowRateFloat
Float-32
R
Floating point value for flow rate
CFM
None
40053-54
Temperature1Float
Float-32
R
Floating point value for Temperature 1
°F
None
40055-56
Temperature2Float
Float-32
R
Floating point value for Temperature 2
°F
None
40057-58
PercentFlowFloat
Float-32
R
Floating point value for Percent Flow
%
None
40059-60
FullScaleFlow
Float-32
R
Floating point value for Full Scale Flow
CFM
None
40061-62
ServoPossition
Float-32
R
Actual position in encoder counts of air plate
Encoder
Counts
40063-64
ServoMotorFault
Float-32
R
Servo Motor Error Code (diagnostic use)
40065-66
ResetServoMotor
Float-32
R/W
1=Reset servo motor
40067-68
AI1_Calibration_Enable Float-32
R/W
0=Read only on Calibration Registers
1=Write enable
40069-70
AI1_Zero
Float-32
R/W
0=No action, 1=Zero Adjust
40071-72
AI1_Cal
Float-32
R/W
0=No action, 1=Cal Adjust
40073-74
AI1_Applied
Float-32
R/W
Actual calibrated value appled to the sensor