29
4
Allow Temp 1 Calibration:
Allows an
administrator to select if the calibration values of
Temp 1
are on
LOCK
(selected in Figure 5.10) or
UNLOCK
. If
UNLOCK
is selected the calibration
values can be changed.
Temperature sensors are not calibrated channels/
sensors. If a temperature is reading incorrectly,
try using a different probe/sensor. If available,
temperature reading may be
verified
with a known
temperature source, temperature dry-block or
thermocouple simulator. If the actual (sensor)
and verified (source) readings are off more than
+/- 2.0°F, the flowbench reading can be adjusted by
entering the difference between the actual (sensor)
and verified (source) readings as the
Offset
value.
Click the current
Offset
value to open a popup
window that allows the administrator to enter the
new value.
5
Allow Temp 2 Calibration
:
Allows an
administrator to select if the calibration values of
Temp 2
are on
LOCK
(selected in Figure 5.10) or
UNLOCK
. If
UNLOCK
is selected the calibration
values can be changed.
Temperature sensors are not calibrated channels/
sensors. If a temperature is reading incorrectly,
try using a different probe/sensor. If available,
temperature reading may be
verified
with a known
temperature source, temperature dry-block or
thermocouple simulator. If the actual (sensor)
and verified (source) readings are off more than
+/- 2.0°F, the flowbench reading can be adjusted by
entering the difference between the actual (sensor)
and verified (source) readings as the
Offset
value. Click the current
Offset
value to open a popup window
to enter the new value.
Figure 5.10: Sensor Calibration
1
2
3
4
5
5.0 Operation