Function Code: 03H Modbus Response (Measurement Parameter)
Logic
Address
R/W
Item
Number
of Bytes
Information
Type
Description of Data
Transmission
Default
Value
Note
0031H
R
Measurement
Status
2
USHORT
0: Hold Status
1
1: Measurement
Status
0032H
R
Sign Byte
6
CHAR
NTU
mg/l
ASCII
Code
FTU
FNU
ppm
mg/l
0035H
R
Turbidity/SS
Measurement
4
FLOAT
Turbidity/SS
Measurement
0037H-
0050H
Factory Reserved
Function Code: 01H Modbus Response (Dispersion Parameter)
Logic
Address
Item
BIT
Description
Default Value
Note
0070H
LO Alarm
1
Contact on
0 (Contact off)
0071H
Hi Alarm
1
Contact on
0 (Contact off)
0072H
mA too high
1
Contact on
0 (Contact off)
0073H
mA too low
1
Contact on
0 (Contact off)
0074H
Factory Reserved
0075H
Turbidity/SS out of range
1
Contact on
0 (Contact off)
0076H
RLY1 Action *
1
Contact on
0 (Contact off)
0077H
RLY2 Action*
1
Contact on
0 (Contact off)
0078H
Clean Action*
1
Contact on
0 (Contact off)
0079H-
008FH
Factory Reserved
52