7.5 Product Adjustment
After entering setup mode, select “Code” and press to make fine adjustments to the
measurement reading. Users are able to make sample reading adjustments based on actual
sample measurement values without retrieving the sensor for calibration. A PDT icon will
display on the main screen above the pH unit if the product is adjusted (see section 6.4).
Enter “Relay 1” Setup
Press to confirm it.
Users may compare the measurement reading
with the reference value. Press or to
tune the adjustment value in the main display in
order to modify the reading of measurement
which will show in the bottom part of display.
Press to confirm it.