6.2 Real-Time Chart Mode
Real-time chart mode shows a dynamic change of the measuring values during the last 3 minutes.
Under setting mode, users are allowed to set corresponding Turbidity/SS measuring range (see
section 7.4) to adjust the resolution of the curve. The smaller the range is being set, the higher
resolution the display is. When the unit enters the setup or calibration mode, the real-time
graphic will be re-updated after returning to measurement mode. When the measured value
exceeds a set range of the upper and lower limit, the graphics will be presented in the upper and
lower limits dotted line. Real-time chart mode display is shown as below. There are also
real-time measurement value, & unit, and temperature value & unit which are displayed in the
bottom of the screen. The timeline in real-time graphic is divided into 12 depicts, each of which
describes a quarter minute (15 seconds).
3 minutes of main measurement
record curve
Main measurement
value and unit
The current
instantaneous value
Upper limit
Lower limit
15s
14