7.11 Analog Output
After entering setup mode, select “Analog” and press to configure analog output settings.
Select 0~20 mA or 4~20 mA current output. Set the upper and lower limit relative to the desired
measuring range. The smaller the range, the greater the current output resolution. When the
measured value exceeds the higher range limit, the current will remain approximately 22 mA
output. When the measured value exceeds the lower range limit, under 0~20 mA mode the
current output will remain 0mA, under 4~20 mA mode the current output will remain
approximately 2 mA. The exceptional output values can be used as an indication for
determination failure. Under HOLD (measurement) status, the current output maintains the last
output value before HOLD status. For the convenience of ensuring the current settings of an
external recorder or of a PLC controller, the current output will be 0/4 mA or 20 mA under the
analog output setup menu.
Press or to
set the range lower
limit relative to 4 mA
output signal.
Press to confirm it.
Press or to
set the range upper limit
relative to 20 mA output
signal.
Press to confirm it.
Press or to
set the range lower
limit relative to 0 mA
output signal.
Enter “Clock” Setup
Press to confirm it.
Press to confirm it.
Press to confirm it.
Press
or to select
for current output range,
0-20 mA or 4-20 mA.
30