8.3.1.2 Turbidity TC-100, TC-500, TC-3000 Electrode Calibration
•
Turbidity calibration each provides 5 preset standard solution values for selection:
TC-100:
80.0, 40.0, 20.0, 10.0, 1.0
TC-500:
400.0, 200.0, 100.0, 50.0, 5.0
TC-3000:
2400, 1200, 600, 300, 30
Press to confirm it.
Press to confirm it.
Press to confirm it.
Press to confirm it.
Press to confirm it.
Press to confirm it.
Press to confirm it.
Press to confirm it.
Press or to
select the first point closest
to the known solution
value.
Press or to
adjust
the
Turbidity
standard solution value.
Rinse the sensor with
distilled water and place
into the first solution for
measurement.
The sensor will display
the current received.
Wait until the value is
stable.
Press to directly
display the result or wait
for the transmitter to
automatically confirm it.
Press or to
select the second point
closest to the known
solution value.
Press to complete
single point calibration.
Press or to
adjust
the
Turbidity
standard solution value.
Rinse the sensor with
distilled water and place
into the second solution for
measurement.
The transmitter will
display the current
received. Wait until the
value is stable.
Press to directly
display the result or wait
for the transmitter to
automatically confirm it.
Press to complete
two-point
calibration
and directly display the
result.
Press to continue
and proceed to multi-
point calibration
44