7.7 Relay 2
Enter setup of Relay 2. Select the item to turn the relay 2 function ON / OFF. If you select to
turn on relay 2, set relay 2 as “High set-point” alarm or “Low set-point” alarm. Set the value of
set-point (SP) and Hysteresis (Hys.). Refer to the graph below for the relationship of the
parameters (for low point alarms).
Enter “Wiper” Setup
Press
or
to select whether to
turn on Relay 2 or
not. If OFF, the
guide menu will go
to setup of Wiper.
Press to confirm it.
Press to confirm it.
Press
or
to select Hi point
or Lo point alarm
of Relay 2.
Press to confirm it.
Press or
to adjust Hysteresis
(Hys.) value
Press to confirm it.
Press to confirm it.
Low Point Alarm Control
off
SP
Hys.
Turbidity/SS
t (sec)
off
on
on
Press or to adjust
Setting Point (SP) value.
Press to confirm it.
Press
or to
select to activate testing of
Relay 2. If turned on, relay
1 will begin testing, and
RELAY 2 indication lamp
will light up.
26