0019H 4
FLOAT
Hys1
0.1pH/10mV
affected by
sign byte
001BH
Relay 2 *
2 USHORT
0: OFF
1
1: AUTO
001CH 2
USHORT
0: Hi
1
1: Lo
001DH 4
FLOAT
SP2
4.00pH/-1000mV
Data
affected by
sign byte
001FH 4
FLOAT
Hys2
0.1pH/10mV
0021H
Backlight
Brightness*
2 USHORT
0: AUTO
2
1: ON
2: OFF
0022H 2
SHORT
2: Highest brightness
0
SHORT
1: high brightness
SHORT
0: Standard
SHORT
-1: Low brightness
SHORT
-2: Lowest brightness
0023H
Backlight
Sensitivity*
2
SHORT
2: Highest Sensitivity
0
SHORT
1: High Sensitivity
SHORT
0: Standard
SHORT
-1: Low Sensitivity
SHORT
-2: Lowest Sensitivity
0024H
Sample average
of measurements
(Digital Filter) *
2 USHORT
1-60
5
0025H-
0030H
Factory reserved
Note 1 : The actions without * sign only support for function code 03H. The actions with * sign support
function code 03H, 06H, 10H. USHORT data range from 0 to 65535, SHORT data range from
-32768 to 32767.
Note 2 :FLOAT is a 32-bit IEEE 754 format. The above table, for an example, is divided into two 16-bit
register data transmission. The back 16-bit register(CC CD) will be transferred first, and then the first
16-bit register (41 C8) will be transferred later. Every 16-bit format is high-bit in the front and
low-bit in the post. For example, the temperature now is 25.1
o
C. The 16-bit of FLOAT
data(Hexadecimal) will show 41 C8 CC CD. The transmission order is CC CD 41 C8. For detail
description, please refer to ch 9.3 Modbus example description.
55