3.5 Connection of transmitter PC-3110 and accessorial transmitter PH-300T
A.
Connect the GLASS point of transmitter PH-300T’s terminal to the electrode coaxial inner.
(Note: Remove the black conductive rubber); connect the REF point of transmitter
PH-300T’s terminal to the electrode coaxial shield.
B.
See the two-wire distributing system and three-wire distributing system in the following
page.
C.
Sign “PT-1000” on transmitter PH-300T’s terminal is the connector for automatic
temperature compensation probe, PT-1000, or applies a fixed temperature compensation
resistance.
D.
The V+ and V- of transmitter PH-300T’s terminal respectively connect to DC12V+ and –
of the controller.
E.
The S+ and S- on transmitter PH-300T’s terminal respectively connect to GLASS and REF
of the controller.
F.
The I+ and I- on transmitter PH-300T’s terminal are output (4-20mA), which can connect
to devices that receive current signals.
(Note: The current output signal of this
transmitter is not insulating, and thus do not directly connect with a PLC!)
Note: Refer to the following table for proper fixed temperature compensation resistance
Temperature
0
℃
5
℃
10
℃
15
℃
20
℃
R value
1000
Ω
1019.25
Ω
1038.5
Ω
1057.75
Ω
1077
Ω
Temperature
25
℃
30
℃
35
℃
40
℃
45
℃
R value
1096.25
Ω
1115.5
Ω
1134.75
Ω
1154
Ω
1173.25
Ω
Temperature
50
℃
55
℃
60
℃
65
℃
70
℃
R value
1192.5
Ω
1211.75
Ω
1231
Ω
1250.25
Ω
1269.5
Ω
Temperature
75
℃
80
℃
85
℃
90
℃
100
℃
R value
1288.75
Ω
1308
Ω
1327.25
Ω
1346.5
Ω
1385
Ω
13