7.5 Multi-Cal
Enter setup of multi-points calibration to set the number of calibration points. The
function is only for pH measurement.
Enter ”Product adjustment” Setup
Press to confirm it.
Press or to select the number of calibration points. There are
1~5 points to choose from. The factory default is 2 points. When the
calibration reaches the number of setting points, the calibration
procedure will automatically be terminated and display the calibration
result.
Press to confirm it.