8.3.7 Any Calibration
The procedure below is two points calibration of Any mode. First, enter the setup of
Multi-points calibration and set the number of calibration point for 2. (See chapter 7.5
Multi-Cal) Then, go to Calibration menu and select “Any” mode. Operate the instrument as
follow procedure diagram. . For multi points calibration, it also needs to set the number of
points in the Multi-Cal setting in advance, and the calibration procedure is the same.
Obtain the pH value of
first point buffer
solution.
Press or
to adjust pH value.
Press to confirm it.
Press to confirm it.
The instrument
measures the mV value
of the buffers. Press
to directly shows the
determination result, or
wait and read for the
result by automatic
display.
Clean the electrode by
distilled water and then
put it into the first
buffer. If under MTC
(Manual Temp.
Compensation)
temperature mode,
press
or
to adjust temperature
value.
Press to confirm it.
Press to confirm it.
Press to confirm it.
Obtain the pH value of
second point buffer
solution.
Press or
to adjust pH value.
Press or to
change the page of
calibration result display
The instrument
measures the mV value
of the buffers. Press
to directly shows the
determination result, or
wait and read for the
result by automatic
display.
Clean the electrode by
distilled water and then put
it into the second buffer.
If under MTC temperature
mode, press
or
to adjust temperature value.
Or press to decide to
simply make a single point
calibration.