The procedure below is two points calibration of Define mode. First, enter the setup of Multi-points
calibration and set the number of calibration point for 2. (See chapter 7.5 Multi-Cal) Then, go to
Calibration menu and select “Define” mode.
Press to confirm it.
Press to confirm it.
Press to confirm it.
Press to confirm it.
Press to confirm it.
Press to confirm it.
The instrument
measures the mV value
of the buffers. Press
to directly shows the
determination result, or
wait and read for the
result by automatic
display.
Press or to
change the page of
calibration result display
Obtain the pH value of
first point buffer
solution.
Obtain the pH value of
second point buffer
solution.
Press or to
select starting of
calibration and execute
preset calibration
Under MTC(Manual
Temp. Compensation)
temperature mode,
Press
or
to select the first point
buffer solution.
Clean the electrode by
distilled water and then
put it into the first
buffer. If under MTC
(Manual Temp.
Compensation)
temperature mode,
press
or
to adjust temperature
value.
Press
or
to select the second
point buffer solution.
The instrument
measures the mV value
of the buffers. Press
to directly shows the
determination result, or
wait and read for the
result by automatic
display.
Clean the electrode by
distilled water and then
put it into the second
buffer. If under MTC
temperature mode, press
or
to adjust
temperature value.
Or press to decide
to simply make a single
point calibration.
50