4.
Insert the Electrode cable
(H)
through
Lower cover of round junction box
(D)
and Cable fixing
gland
(C)
, and use Lower cover of round junction box
(D)
to fix PP Electrode Protective
Housing
(G)
tightly.
5.
Prepare 15cm cable in the PP pipe, and then fix Cable fixing gland MG16A
(C)
tightly. Leave
Electrode cable
(H)
for about 12-14cm, and split it carefully.
6.
Fix the terminal of electrode coaxial inner on terminal block 1 of round holder; Fix the
terminal of electrode coaxial shield on terminal block 3. (See the instruction of junction box)
7.
Extend the cable to pass through Cable fixing gland
(E)
on Lower cover of round junction
box
(D)
, and fix Cable fixing gland MG16A
(E)
tightly, leaving 12-14cm in the box for split.
8.
Extend the lead coaxial inner and electrode coaxial inner to connect them; extend the lead
coaxial shield to fix on the terminal block 3. Tight up
Upper cover of round junction box
(A)
to finish the installation.
Installation of holder support base
The L-shaped electrode holder support base is installed by finding a appropriate position in
the edge of a pool according to the field’s needs with nails or expansion screws.
1.
Fix the holder fixed plate(C) into the support base (D)
2.
Fix the U-shaped clamp (A) into immersive holder (B)
3.
Combine the item 1. and item 2., tighten it up by the plastic screws (E)
A------Holder U-shaped clamp
B------Immersive holder(PP-100A)
C------Holder fixed plate
D------Support base
E------Plastic screws
8