0024
R/W
Sample
average of
measurements
(Digital
Filter)*
2 USHORT
1-60
20
0025
R/W
Relay 1 type*
2
USHORT
0
:
Free Cl
2
0
1
:
pH
0026
R/W
Relay 2 type*
2
USHORT
0
:
Free Cl
2
0
1
:
pH
0027
R/W
Cln/Auto Zero
type*
2 USHORT
0
:
Clean
0
1
:
Auto Zero
0028
R/W
Initial Volt*
2
USHORT
0 ~1000 mV
100mV
Absolute
value
0029
R/W
Gain* 2
USHORT
0
~100mV/
μ
A 10mV/
μ
A
Absolute
value
002A
R/W
Motor* 2
USHORT
0
:
OFF
1
1
:
ON
002B
R/W
Water Detect*
2
USHORT
0
:
OFF
1
1
:
ON
002C
R/W
Remote zero
point*
2 USHORT
0
:
OFF
0
2:Error
(Read Only)
1
:
ON
2
:
Error
002D
R/W
Remote Span
calibration*
2 USHORT
0
:
OFF
0
2:Error
(Read Only)
1
:
ON
2
:
Error
002E
R/W
Remote Span
input*
4 FLOAT
Span
adjustment
0~3
mg/l
0030 Factory
reserved
Note 1 : The actions without * sign only support for function code 03H. The actions with * sign
support function code 03H, 06H, 10H.
Note 2 :FLOAT is a 32-bit IEEE 754 format. The above table, for an example, is divided into two
16-bit register data transmission. The back 16-bit register(CC CD) will be transferred first, and
then the first 16-bit register (41 C8) will be transferred later. Every 16-bit format is high-bit in
the front and low-bit in the post. For example, the temperature now is 25.1
o
C. The 16-bit of
FLOAT data(Hexadecimal) will show 41 C8 CC CD. The transmission order is CC CD 41 C8.
For detail description, please refer to Ch 9.3 Modbus example description.
Note3 : USHORT means unsigned short integer.