7.10 Motor
Enter setup of motor. Select to turn on or turn off the motor operation(can also be set in the
calibration mode). Under measurement mode, the motor must be kept running. If the motor stop
running, it will affect the correctness of chlorine measurement. At the time, there is a “M-OFF”
warming symbol on the top right corner of the display. In addition, when the water level detection
function is turned on, under measurement mode if electrode circuit open or water outage(the water
level drop to the bottom of the silver electrode) happens, or if the measurement is kept constantly at
near zero, then the instrument will judge it and stop the motor automatically after a certain of time. It
prevents residual chlorine electrode continuously rub against without water sample and result to
damage. It also shows “M-OFF” warning symbol on the top right corner of the display and goes into
HOLD status. When turning off the water detection function, under measurement mode, regardless of
the electrode circuit open, water outage, or constant zero measurement, the motor will keep running.
Enter
“
Relay 1
”
Setup
Press to confirm it.
Press to confirm it.
Press to confirm it.
Press or to
activate the motor or not.
Press or to turn on
or turn off water level detection
function for auto motor stop.
35