7.12 Relay 2
Enter setup of Relay 2. Select the item to turn on or turn off the relay 2 function. If you select to turn
on the relay 1, then select for using relay 1 as Free Cl
2
or pH relay function, and then select “High
set-point” alarm or “Low set-point” alarm. Set the value of set-point (SP) and Hysteresis (Hys.). The
relationship between parameters can refer to an explanatory diagram of the box (as a high point
alarm).
If select relay 2 as Free Cl
2
control, the procedure is as follow:
Low Point Alarm Control
off
SP
Hys.
Free Cl
2
(mg/l)
t (sec)
off
on
on
Enter “Clean/Auto Zero” Setup
Press to confirm it.
Press to confirm it.
Press to confirm it.
Press to confirm it.
Press to confirm it.
Press to confirm it.
Press to confirm it.
Press
or
to select to activate
REL2 or not. If
not, the guide
menu goes to setup
of Cln/Auto Zero.
Press
or
to select use REL2
as Hi point or Lo
point alarm.
Press
or to
select to activate a test of
REL2 or not. If “ON”, the
relay 2 is in action, and
the “ACT” indication
light will light up.
Press or
to adjust Hysteresis
(Hys.) value
Press or
to adjust set-point
(SP) value
Press
or
to select Relay 2
for Free Cl
2
measurement.
38