1. Specifications
1.1 Specifications
Main specifications:
CT-6110-POL (Non-reagent type free chlorine transmitter)
Measuring mode
Residual(Free) Chlorine & pH / Temperature, Simultaneous display
Measuring
range
Free Cl
2
0.00~3.00mg/l (Free Cl
2
corresponding range adjustable for graphical display)
pH -2.00~16.00pH
Temp 0.0~60.0
C
Resolution
Free Cl
2
0.01mg/l
pH 0.01pH
Temp 0.1
C
pH compensation/
accuracy
pH sensor auto compensation/ manual compensation/ OFF / ±0.01pH(±1 digit)
Temp. compensation/
accuracy
NTC-30K(compensation range: 0~45
C)/ ±0.2
C (±1 digit)
Display
Large LCM with sensitization sensor for auto/manual illumination function and
contract function
Display mode
Text mode: Numerical display
Chart mode: 3 mins real-time dynamic graph
(Free Cl
2
corresponding range adjustable)
Trace mode: Set up from 3 mins to 4 weeks duration of the measured value trend graph
(Free Cl
2
corresponding range adjustable)
Analog output
Isolated DC 0/4~20mA corresponding to Free Cl
2
or pH, max. load 500
Voltage output
DC ± 12V, 1W max. for PH-300T(Optional)
Logbook
50 event records
Serial interface
Isolated RS-485(Modbus RTU or ASCII)
Setting
Contact
240VAC, 0.5A Max. (recommended)
Activate
Two individual Free Cl
2
or pH, Hi/Lo selectable, limited programmable
Clean/Auto Zero
240VAC, 0.5A Max. (recommended),
ON:0~99 min. 59 sec. OFF:0~999 hours 59 min.
Cl
2
calibrati
on
Zero point
Electrode open circuit
or apply activated charcoal filtered water (residual chlorine free water) / pure water
Span
Sampling DPD measurement input calibration or standard buffer calibration
pH calibration
Any, Tech & NIST buffer calibration
Diagnostic function
Temperature sensor, zero-point & slope diagnosis, water/power outage,
logbook record or alarm symbol
Measurement system
Measuring principle
Polarographic (No reagent needed for measurement)
Residual chlorine
electrode
2 in 1 rotating electrode (Working electrode: gold, Counter electrode: silver)
With relative slope index for electrode aging determination
Temperature probe
NTC-30K
pH electrode
Optional(No need for pH compensation when pH range within pH 6.5~7.5)
Cl
2
electrode cleaning
Ceramic beads auto cleaning
Motor protection
Auto over heat protection
Water outage/
Electrode open
diagnosis
When water outage or electrode open, after auto diagnosis the motor auto shuts down
4