7.7 Polarogram
Enter setup of polarogram. Set the scan voltage of initial volt, final volt, scan rate, max. value
and min. value of the diagram. After setting, it starts the scanning procedure and shows the
scanning curve. Users may get the scanning voltage and current relationship from the
polarogram. Select the voltage of stable current range(which means voltage change but little
change in current range), and then set the voltage as initial volt in the transmitter’s polar. volt
setting (Please refer to ch7.6 Polar. Volt.).
Enter “pH function”Setup
Start the scanning
procedure and display the
scanning polarogram.
Press to confirm it.
Press to confirm it.
Press to confirm it.
Press to confirm it.
Press to confirm it.
Press to confirm it.
Press to confirm it.
Select YES and press to
confirm it.
Select NO to remain
the setting and reserve
the record, press
to confirm it.
Press or to
input initial volt.
Press or to
input final volt.
Press or to
input the scanning rate.
Press or to
input the maximum
display value.
Press or to
input the minimum
display value.