Motor
Motor setting, to activate the motor and water sample level
detection function
Relay 1
First relay setting, to choose action off or Hi/Lo alarm
Relay 2
Second relay setting, to choose action off or Hi/Lo alarm
Cln/Auto Zero
Automatic wash/auto zero time setting, to set clean device or
zero point filtered water device ON and OFF duration
Analog
Current output according to Free Cl
2
or pH setting range
Clock
Clock setting
(When out of power and reboot it, the
instrument’s time setting can maintain to the real time. If
not, please replace the inner 3V CR2025 battery.)
RS-485
RS485 serial interface (Modbus protocol)
Black-light
Backlight setting, to set Auto/ON/OFF backlight, brightness,
and sensitivity
Contrast
Contrast of screen setting
Digital Filter
Take every serial 1~60 measurements, average them
continuously, and make it as the readings
Logbook
Event recorder logbook (50 data)
Return
Setting of returning to the measurement mode
Code
Security code of set-up mode. The set-up code is precedential to
calibration code, thus it can pass a different security code of
calibration.
Language
Available for English, Traditional Chinese, Simplified Chinese