Form No. 702718 Rev. A March, 2005
87
Owner’s Manual - Regal Intensive Care Unit
New Style Units
1.
Do not remove the oxygen sensor from the partiion wall.
2.
With the oxygen sensor in place in the partition wall,
unscrew the end cap from the oxygen sensor (Figure 96).
3.
On the fuel cell, remove the shorting button - refer to the
instructions packed with the fuel cell.
Note:
Watch the oxygen concentration meter (Figure 38) as you
perform
Step 4
. The black calibration needle moves to the right,
then settles down.
4.
Place the fuel cell into the oxygen sensor body. The end
of the fuel cell with the metal disk goes in first.
5.
Gently, screw the sensor end cap (Figure 96) back on. Be
careful, the threads are very fine and are easily damaged.
6.
Proceed to
Finish the Installation
below.
Finish the Installation
1.
Close and secure the electronics compartment door.
2.
Close and secure the filter/fans compartment.
3.
Using the calibration knob (Figure 98), position the black
needle in the oxygen concentration meter to 20.5% (the
percentage of oxygen in atmospheric air).
4.
When you are finished working with the fuel cell, wash
your hands thoroughly.
Summary of Contents for REGAL INTENSIVE CARE UNIT 12019-01-NGFREP
Page 12: ...x SSCI Suburban Surgical Company Inc Table of Contents Notes...
Page 20: ...8 SSCI Suburban Surgical Company Inc Chapter 1 General Information Notes...
Page 44: ...32 SSCI Suburban Surgical Company Inc Chapter 2 Description of Components Notes...
Page 148: ...136 SSCI Suburban Surgical Company Inc Chapter 5 Repairs Replacements Notes...
Page 168: ...156 SSCI Suburban Surgical Company Inc Chapter 6 Troubleshooting Notes...
Page 170: ...A 2 SSCI Suburban Surgical Company Inc Appendix Thispageleftblankintentionally...
Page 174: ...A 6 SSCI Suburban Surgical Company Inc Appendix Thispageleftblankintentionally...
Page 183: ...Inside back cover This page left blank intentionally...