142
SSCI - Suburban Surgical Company, Inc.
Chapter 6 - Troubleshooting
Oxygen flow is excessive.
First:
Make sure your oxygen pressure regulator is correctly
adjusted?
Second:
Is the oxygen flow meter adjusted properly? Refer to
Controlling Oxygen Flow Rate
on
Page 56.
Third:
Is the set point on the oxygen concentration meter
correctly set? Refer to
Using Oxygen
on
Page 51.
Fourth:
Your solenoid valve may be defective. Order P/N
853440 and refer to
Page 118
for replacement instructions.
Fifth:
Your oxygen flow meter may be defective. Order
P/N 853456 and refer to
Page 120
for replacement instructions.
Remedial Action