Instructions (continued)
To Select a Procedure Configuration:
Procedure Configurations Table
NOTES:
• The multi-lesion capability allows the generator to lesion tissue using multiple electrodes. The RF energy is provided in short bursts that are timed to allow all
the electrodes to heat at the same time. RF energy and voltages for stimulation are never supplied concurrently. Voltages used for stimulation may only be
applied to one electrode at a time.
• RF energy is delivered through two electrode connector ports, each of which is segmented into two logical ports providing a total of four logical ports. The
generator can support several combinations of electrode configurations. These combinations may be mixed between the two electrode connector ports.
• DO NOT use configurations other than those described below. Failure to comply will result in a system error.
PROCEDURE
TYPE
NUMBER OF
ELECTRODE(S)
GENERATOR CONNECTOR 1-2
GENERATOR CONNECTOR 3-4
GROUND PAD
CONNECTOR
FIGURE
NUMBER
Monopolar Procedure Configurations:
capability allows the generator to create a lesion using a monopolar electrode with a ground pad.
Single Monopolar
1
MultiGen Cable (0406-900-100)
Empty
Ground Pad Cable
(0406-850-200)
7
Empty
MultiGen Cable (0406-900-100)
Two Concurrent
Monopolar
2
MultiGen Cable (0406-900-100)
Empty
Ground Pad Cable
(0406-850-200)
7
Empty
MultiGen Cable (0406-900-100)
Three Concurrent
Monopolar
3
MultiGen Cable (0406-900-100)
MultiGen Cable (0406-900-100)
Ground Pad Cable
(0406-850-200)
7 and 8
MultiGen Cable 0406-900-100)
MultiGen Cable (0406-900-100)
Four Concurrent
Single Monopolar
4
MultiGen Cable (0406-900-100)
MultiGen Cable (0406-900-100)
Ground Pad Cable
(0406-850-200)
7
Parallel Bipolar Procedures Configurations:
capability allows the generator to create a lesion using two monopolar electrodes. The generator monitors the
temperatures of both electrodes and controls the higher of the two electrode temperatures. Both temperatures are displayed on the screen.
Single Parallel
Bipolar
2
MultiGen Cable (0406-900-100)
Empty
Empty
9
Empty
MultiGen Cable (0406-900-100)
Two Concurrent
Bipolar
4
MultiGen Cable (0406-900-100)
MultiGen Cable (0406-900-100)
Empty
Monopolar and Parallel Bipolar Combination Procedure Configurations
Monopolar and a
Parallel Bipolar
3
MultiGen Cable (0406-900-100)
MultiGen Cable (0406-900-100)
Ground Pad Cable
(0406-850-200)
7 and 9
MultiGen Cable (0406-900-100)
MultiGen Cable (0406-900-100)
Two Concurrent
Monopolar and a
Parallel Bipolar
4
MultiGen Cable (0406-900-100)
MultiGen Cable (0406-900-100)
Ground Pad Cable
(0406-850-200)
7