46
www.stryker.com
EN
0406-900-702 Rev-E
Specifications (continued)
Lesion Time:
0 to 999 seconds; default adjustable at system and file level
RF Frequency:
Conventional Heat Lesion:
Pulse Mode
1
:
1 MHz, ac 10%, DualWave Waveform: aperiodic damped sinusoid (see
Appendix D - DualWave Waveform
section)
1 MHz sine wave with selected pulse frequency and on time, ac 10%
Stimulation Frequency:
1 to 200 Hz, separate defaults for sensory and motor stimulation adjustable at
system and file level
Stimulation Type:
Biphasic square wave, voltage or current regulated; no net DC
Stimulation Pulse Width:
0.1 ms<Pulse Width< 5 ms, ac 10%, separate default for sensory and motor
stimulation adjustable at system and file level
Stimulation Amplitude Voltage Regulation Mode:
0-10 V peak; ac 10%, current limited to 20 mA
Volume Adjustment:
0 dBA to 65 dBA; exceptions include alarms and lesion creation indication
Printer Compatibility:
PCL3 compliant or a more current print driver
Environmental Conditions:
Operation
Storage and Transportation
Temperature:
30 °C
10 °C
65 °C
-34 °C
Relative Humidity:
80 %
15 %
Atmospheric Pressure:
1
Pulse mode frequency and width tolerances may vary based on the number of active electrodes and their corresponding pulse mode settings.
Guidance and manufacturer's declaration - electromagnetic emissions
The RF MultiGen (generator) REF 0406-900-000 is intended for use in the electromagnetic environment specified below. The customer or the user of the RF
MultiGen (generator) REF 0406-900-000 should assure that it is used in such an environment.
Emissions test
Compliance
Electromagnetic environment - guidance
RF emissions
CISPR 11
Group 1
The RF MultiGen (generator) REF 0406-900-000 uses RF energy only for its internal
function. Therefore, its RF emissions are very low and are not likely to cause any
interference in nearby electronic equipment.
RF emissions
CISPR 11
Class A
The RF MultiGen (generator) REF 0406-900-000 is suitable for use in all establishments
other than domestic establishments and those directly connected to the public low-voltage
power supply network that supplies buildings used for domestic purposes.
Harmonic emissions
IEC 61000-3-2
Class A
Voltage fluctuations/flicker
emissions
IEC 61000-3-3
Complies