11. Using rubbing alcohol, lubricate the two line manifold fitting and the non-AST air cell fitting.
12. Apply firm, constant force while you turn the air cell fitting back and forth to connect the non-AST air cell to the two line
manifold.
13. Install a nylon cable tie onto the two line manifold/non-AST air cell fitting, 0.25 inches from the end of the hose.
14. Inflate and verify that the air cell inflates.
15. Button both sides of the non-AST air cell to the side bolsters.
16. Move the fire barrier over the air cells toward the foot end to install the fire barrier.
17. Zip the fire barrier.
18. Zip the support surface top cover onto the bottom cover.
19. Run the pump diagnostic test. All tests must pass.
20. Verify proper operation before you return the product to service.
A
AS
ST
T a
aiirr cce
ellll rre
ep
plla
acce
em
me
en
ntt
T
To
oo
ollss rre
eq
qu
uiirre
ed
d::
•
Diagonal pliers
•
Rubbing alcohol
P
Prro
occe
ed
du
urre
e::
1. Unzip and remove the support surface top cover.
2. Unclip the head end and foot end center belt retainers from the bottom cover.
3. Unzip the fire barrier.
4. Move the fire barrier over the side bolsters and air cells toward the head end to remove the fire barrier.
5. Using diagonal pliers, cut the four nylon cable ties (A) from the two line manifold/AST air cell fittings (B) (Figure 6).
6. Unbutton both sides of the AST air cell from the side bolsters. Repeat for all four AST air cells.
7. Slide the AST air cell through the center belt (C) and remove. Repeat for all four AST air cells.
B
A
C
D
F
Fiig
gu
urre
e 6
6 –– A
AS
ST
T a
aiirr cce
ellllss rre
em
mo
ovve
ed
d
8. Feed the AST cable plug through the bottom cover sleeve. Slide the sleeve toward the support surface to access the
AST cable spiral wrap.
N
No
otte
e -- Make note of how the cable spiral wrap is secured.
9. Starting at the support surface side, unwrap the cable spiral wrap to remove the AST cable (D).
N
No
otte
e -- When you install the AST air cells, the distance from the edge of the CPR connector to the AST connector should
be a minimum of 5.90 inches. There should be 3.15 inches from the end of the cable spiral wrap to the end (Figure 7).
2941-009-002 Rev B.0
17
EN