G -- Always plug the power cord into the bed auxiliary outlet or a grounded power respectable. Do not use an
extension cord.
1. Plug the support surface hoses into the pump. Remove all weight from the support surface.
2. Power on the pump.
3. As the software loads, press the O
ON
N button within five seconds (Figure 1), press, and hold the A
Accttiio
on
n button next to the
H
Ho
om
me
e icon (Figure 2).
IsoAir
F
Fiig
gu
urre
e 1
1 –– S
So
offttw
wa
arre
e llo
oa
ad
diin
ng
g
F
Fiig
gu
urre
e 2
2 –– S
Se
errvviicce
e ssccrre
ee
en
n h
ho
om
me
e
4. Press the A
Accttiio
on
n button next to the S
Se
errvviicce
e icon to start the diagnostic mode (Figure 3).
F
Fiig
gu
urre
e 3
3 –– S
Se
errvviicce
e ssccrre
ee
en
n e
en
nttrryy
IsoAir
F
Fiig
gu
urre
e 4
4 –– D
Diia
ag
gn
no
ossttiicc m
mo
od
de
e a
an
nd
d sso
offttw
wa
arre
e vve
errssiio
on
n
N
No
otte
e -- The software version is located on the service screen (Figure 4).
5. Review the diagnostic result.
a. If the diagnostic result is OK, the test passes.
b. If the diagnostic result is not OK, investigate the diagnostic test result.
P
Prro
otte
eccttiin
ng
g a
ag
ga
aiin
nsstt e
elle
eccttrro
osstta
attiicc d
diisscch
ha
arrg
ge
e ((E
ES
SD
D))
C
CA
AU
UT
TIIO
ON
N
•
Always use ESD protective equipment before you open antistatic bags and service electronic parts.
•
Do not place unprotected circuit boards on the floor.
N
No
otte
e -- Always ship the circuit boards back to Stryker. Use the antistatic bag that the new board was originally shipped in.
2941-009-002 Rev B.0