W
Wa
arrn
niin
ng
g//C
Ca
au
uttiio
on
n//N
No
otte
e D
De
effiin
niittiio
on
n
The words W
WA
AR
RN
NIIN
NG
G, C
CA
AU
UT
TIIO
ON
N, and N
NO
OT
TE
E carry special meanings and should be carefully reviewed.
W
WA
AR
RN
NIIN
NG
G
Alerts the reader about a situation which, if not avoided, could result in death or serious injury. It may also describe
potential serious adverse reactions and safety hazards.
C
CA
AU
UT
TIIO
ON
N
Alerts the reader of a potentially hazardous situation which, if not avoided, may result in minor or moderate injury to the
user or patient or damage to the product or other property. This includes special care necessary for the safe and effective
use of the device and the care necessary to avoid damage to a device that may occur as a result of use or misuse.
N
No
otte
e -- Provides special information to make maintenance easier or important instructions clearer.
S
Su
um
mm
ma
arryy o
off ssa
affe
ettyy p
prre
ecca
au
uttiio
on
nss
Always read and strictly follow the warnings and cautions listed on this page. Service only by qualified personnel.
W
WA
AR
RN
NIIN
NG
G
•
Always plug the power cord into the bed auxiliary outlet or a grounded power respectable. Do not use an extension cord.
•
The use of accessories, transducers, and cables, other than those specified or provided by the manufacturer, could
result in increased electromagnetic emissions or decreased electromagnetic immunity and result in improper operation.
•
Portable RF communications equipment, including peripherals such as antenna cables and external antennas, should
be no closer than 12 inches (30 cm) to any part of the IIsso
oA
Aiirr system, that includes cables specified by the manufacturer.
•
Avoid stacking or placing equipment adjacent with other equipment to prevent improper operation of the products. If
such use is necessary, carefully observe stacked or adjacent equipment to make sure that they are operating properly.
•
This equipment/system is intended for use by healthcare professionals only. This equipment/system may cause radio
interference or may disrupt the operation of nearby equipment. It may be necessary to take mitigation measures, such
as reorienting or relocating the IIsso
oA
Aiirr system or shielding the location.
C
CA
AU
UT
TIIO
ON
N
•
Improper usage of the product can cause injury to the patient or operator. Operate the product only as described in this
manual.
•
Do not modify the product or any components of the product. Modifying the product can cause unpredictable operation
resulting in injury to patient or operator. Modifying the product also voids its warranty.
•
Do not tie the support surface to the bed frame with the bottom center D-ring straps to avoid the risk of equipment
damage. The bottom center-D ring straps are for storage only.
•
Always use ESD protective equipment before you open antistatic bags and service electronic parts.
•
Do not place unprotected circuit boards on the floor.
2941-009-002 Rev B.0
3
EN
Summary of Contents for 2941-000-001
Page 2: ......