P
Prro
occe
ed
du
urre
e::
1. Unplug the power cord from the wall outlet.
2. Unplug the support surface tubing from the pump.
3. Place the pump face down on a work surface.
4. Using a T20 Torx driver, remove the six screws that secure the back pump housing to the front pump housing. Save the
screws.
5. Rotate the pump assembly to the right, so the front of the pump is on your left.
6. Separate the front and back pump housing. Place the back of the pump down on the work surface.
7. Make sure you use ESD protection. See
Protecting against electrostatic discharge (ESD)
8. Unplug the LED board cable from the PCBA assembly.
N
No
otte
e -- Make note of the LED board cable connection location for reinstallation.
A
B
B
F
Fiig
gu
urre
e 2
22
2 –– L
LE
ED
D b
bo
oa
arrd
d
9. Using a stubby #2 Phillips screwdriver, remove the four screws (B) that secure the LED board assembly (A) to the front
pump housing assembly. Save the screws.
10. Unplug the cable from the LED board assembly (A). Save the cable.
11. Remove and discard the LED board assembly.
12. Reverse steps to reinstall.
13. Plug the pump cord into an accessible outlet.
14. Run the pump diagnostic test. All tests must pass.
15. Verify proper operation before you return the product to service.
P
Pu
um
mp
p rre
ep
plla
acce
em
me
en
ntt
T
To
oo
ollss rre
eq
qu
uiirre
ed
d::
•
T20 Torx driver
•
Wire cutters
•
#2 Phillips screwdriver
•
ESD system
P
Prro
occe
ed
du
urre
e::
1. Unplug the power cord from the wall outlet.
EN
36
2941-009-002 Rev B.0