50
Maintenance
Follow the instructions in this section for proper care of the device.
Note:
Repairs and equipment modifications shall be performed only by
Stryker-authorized personnel. Stryker Endoscopy assumes no product liability
or warranty responsibility for devices repaired by or purchased from third-
party service organizations.
Inspecting the Console
Inspect the console regularly for cleanliness. If the console appears dirty,
repeat the above cleaning and disinfection procedure.
Regularly inspect the lock ring on the front of the console (located at the
camera connection port). Confirm the lock ring is not loose; tighten if
necessary by manually turning the lock ring clockwise.
Inspect the console before each use. If a problem listed below is observed or
suspected, contact your Stryker representative or return the device to Stryker
for service.
•
If the console is unresponsive to touchscreen button presses or cannot
power on
•
Visible cuts or breaks in any cables
•
Unacceptable deterioration such as (but not limited to) corrosion,
discoloration, pitting, cracked seals, or abnormal noises
Inspecting the Camera Head
Inspect all components of the camera head for cleanliness before each use. If
fluid or tissue buildup is present, repeat the above cleaning and sterilization
procedures.
Inspect the camera head before each use. If a problem listed below is
observed or suspected, contact your Stryker representative or return the
device to Stryker for service
•
If the image output performance of the camera is unacceptable. (For
camera heads 1688210105 and 1688710105, use a coupler that has been
inspected per the coupler user guide.) The camera head shall be able to
provide an image that is clear and focusable with adequate response to
lighting of various scenes.
•
If the console is unresponsive to pressing camera head buttons
•
Visible cuts or breaks in the camera head cable or keypad area
•
Unacceptable deterioration such as corrosion, pitting, cracked seals, or
abnormal noises