25
Follow the instructions below to perform the White Balance test:
1. Ensure that a scope, camera head, light source, and display monitor are
connected to the camera system, and that the camera console, light
source and display monitor are powered on.
2. Point the scope tip at several stacked white gauze pads, a white
laparoscopic sponge, or any clean white surface.
3. Look at the display monitor and make sure there is no visible glare off of
the white surface of the image.
4. Press the WB button on the Home screen (or press and quickly release a
camera head button, if it has been programmed for White Balance) until
“WHITE BALANCE IN PROGRESS” appears on the display monitor.
5. Continue pointing the scope at the white surface until “WHITE BALANCE
COMPLETE” appears on the display monitor. The image may change color.
If you cannot achieve an acceptable White Balance, refer to the
Troubleshooting section.
Note:
The White Balance test is not available when SPY mode is on.
Controlling Remote Video Accessories
When connected with the provided remote cables, the camera can remotely
control up to two functions of a video accessory such as a Stryker digital
capture console. Commonly this enables the user to capture images or start
and stop video recording.
Remote video accessories can be controlled with the camera head’s Camera
button or the console touchscreen. See the following sections, Using the
Touchscreen Interface and Using the Camera Head Buttons.
See the Setup section for instructions about connecting a video accessory to
the console’s Remote outputs.