28
SPY Screen
SPY mode allows the camera to visualize near-infrared light produced by the
L11 LED Light Source (0220230300). The SPY screen allows the user to adjust
image settings within SPY mode.
Note:
White Light must be activated on the L11 LED Light Source as a
preliminary step to activate SPY mode.
See the SPY Mode Requirements section for required equipment and
conditions to enable SPY mode.
2
Before using SPY mode, be familiar with the Warnings: SPY
Mode section of this user manual and all warnings and
instructions in the light source user manual. Failure to follow all
warnings can result in severe eye injury to the patient or user.
1. SPY Mode selection:
Press and briefly hold the arrows to scroll through
the following SPY modes. The icon shown next to each SPY mode appears
in the top-left corner of the display monitor when the mode is active.
(red, green, and blue)
•
Overlay:
The console outputs a white light image
with ICG fluorescence indicated by the color green.
(all grey)
•
Contrast:
The console outputs an image with ICG
fluorescence indicated by the color white. All other
areas of the image appear dark.
(all green)
•
ENV:
The console outputs a greyscale white light
image with ICG fluorescence indicated by the color
green.