Sterilization
After performing the cleaning instructions specified above, perform one
of the following sterilization cycles.
Note:
For necessary materials and
equipment, see the Materials and Equipment table.
Sterrad
1. Clean and prepare the device as recommended in this user manual.
Ensure the soaking cap is installed.
2. If using a sterilization tray (optional), follow any additional instructions
provided with the tray. Use only trays that are compatible with Sterrad.
3. Double wrap the device (or tray) prior to sterilization.
4. Sterilize the device using one of the following Sterrad sterilization
systems:
•
100S (Short or Long
1
cycle)
•
NX (Standard or Advanced cycle)
•
100NX (Standard or DUO cycle)
•
NX AllClear (Standard or Advanced cycle)
•
100NX AllClear (Standard or DUO cycle)
1
Sterrad 100S Long cycle is intended for use outside the U.S. only.
5. Allow the device to cool to room temperature before reconnecting it to
the camera system. Otherwise, the lens can fog during use or the console
can produce a system error.
Steris/Amsco V-PRO
1. Clean and prepare the device as recommended in this user manual.
Ensure the soaking cap is installed.
2. If using a sterilization tray (optional), follow any additional instructions
provided with the tray. Use only trays that are approved for sterilization
with V-PRO.
3. Double wrap the device (or tray) prior to sterilization.
4. Sterilize the device using one of the following V-PRO sterilization systems:
•
V-PRO 1 (Standard cycle)
•
V-PRO 1 Plus (Non-Lumen or Lumen cycle)
•
V-PRO 60 (Non-Lumen, Lumen, or Flexible cycle)
•
V-PRO maX (Non-Lumen, Lumen, or Flexible cycle)
•
V-PRO maX 2 (Non-Lumen, Lumen, Flexible, or Fast Non-Lumen
1
cycle)
1
Warning:
When using the Fast Non-Lumen cycle, follow the instructions provided with the V-PRO maX 2 system
regarding use of rigid containers or sterilization wrap. The cycle is intended to be used with sterilization pouches. Also
observe the Fast Non-Lumen cycle weight limit of up to 11 pounds (4.99 kg) of instruments.
5. Allow the device to cool to room temperature before reconnecting it to
the camera system. Otherwise, the lens can fog during use or the console
can produce a system error.