21
2. Attach an endoscope to the coupler.
Before each use, check the outer surface of the endoscope to
ensure there are no rough surfaces, sharp edges, or protrusions.
For 1688210105, 1688610122, and 1688710105:
Note:
For a list of endoscopes that are compatible with SPY mode,
see the user manual for the L11 LED Light Source.
3
1
2
•
Depress the endobody clamp
1
and insert an endoscope into
the endobody
2
.
•
Release the endobody clamp to secure the endoscope.
For 1688310130:
1
2
3
•
Lock the endobody brake
1
by rotating it clockwise (when
facing the camera head buttons).
•
Twist the endobody clamp
2
as shown, and hold it open.
•
Insert the endoscope into the endobody clamp.
•
Release the endobody clamp. It will return to the original
position and hold the endoscope.
•
If the endoscope is not secured in a fixed position, slightly
rotate the endobody clamp in either direction until the lock
engages with an audible click.
3. Attach a light cable from the light source to the light post on the
endoscope
3
.
Summary of Contents for 1688 Series
Page 2: ......
Page 22: ...18 Wiring Diagram 0240200100 1688010000 0220230300 0240 031 050 3 4 2 6 1 5 ...