SYMPTOM
PROBABLE CAUSE
ACTION
The rover canisters are
not clean after a cleaning
cycle.
The water temperature
supplied to the docker is
not adjusted for optimal
cleaning.
For optimal cleaning of the canisters, use a water temperature
of 37.8 to 43.3 °C [100 to 110 °F]. See the
instructions for use
supplied with the docker.
The bottle of detergent is
empty.
Replace the bottle of the detergent. See
Section 5.3 For Use
With
and the
instructions for use
supplied with the detergent.
An error occurs while the
rover is docked and the
rover cannot be removed
from the docker.
The fluid connectors
prevent the removal of the
rover from the docker.
Remove power from the docker; apply power to the docker. Wait
20 seconds.
Remove the rover from the docker. Cycle docker power. Wait 60
seconds; try docking the rover again.
If the problem persists, the
fluid connectors may be
damaged.
Contact Stryker Neptune Customer Service. See
Section 1.3
Contact Information
.
The rover cannot be
removed from docker.
The two rear casters are
locked.
Make sure the two rear casters are unlocked. DO NOT lock the
casters when the rover is docked. Pull the rover away from the
docker.
The operator failed to
acknowledge the docking
cycle is complete.
From the DOCKING MODE screen, touch the RELEASE FROM
DOCK(er) button to disconnect the rover from the docker.