0703-001-700 Rev-C
EN
www.stryker.com
33
8.2.2 To Install Disposable Components
NOTES:
▪ The fluid suction tubing and suction accessory are
applied parts. See
Section 17 Glossary
.
▪ If the manifold is valid and installed correctly, a ‘task
complete’ audible indicator will occur, the manifold
suction indicator will illuminate white or pink, and the
START SUCTION button will be enabled.
▪ See
Section 17 Glossary
for the definitions of a new,
valid, used, expired, and invalid manifold.
8.2.2.1 To Prepare for Fluid Suction
1. While aligning the tabs, insert a new (valid) manifold(s)
into the manifold receptacle(s) (Figure 26).
2. Fully rotate the manifold(s) clockwise to lock into the
manifold receptacles. See the
instructions for use
supplied with the manifold.
3. Attach the fluid suction tubing to the port(s) of the
installed manifold(s). ALWAYS close unused manifold
ports.
4. Attach a fluid suction accessory to the end of the
suction tubing, if desired.
Figure 26 – To Prepare for Fluid Suction
5. From the control screen, touch the RESET VOLUME
button to reset the fluid volume value of each canister
to zero milliliters, if required. A CONFIRM dialog will
appear (Figure 27) if the button is selected. Touch the
CONFIRM button to make the change.
Suction OFF
Reset 4L and 20L
canister volumes?
Volume counters will
reset to zero.
Cancel
Confirm
Reset Volumes
Figure 27 – Reset Volumes Dialog
0000160552, Rev. C Effective Date: Nov 25, 2015 1:43:15 PM
Print Date: Nov 25, 2015 02:19:13 PM