Figure 39 – To Dispose of Suction Components
8.4.1.2 To Remove Evacuator Components
1. From the control screen, touch the EVACUATE SMOKE
button. From the EVACUATE SMOKE dialog, touch the
OFF button to stop smoke evacuation.
2. Gather the smoke evacuator tubing with any
connected attachments toward the smoke evacuator
filter. Remove the tubing and any connected
attachment from the smoke evacuator filter (Figure
40).
3. Properly dispose of the used smoke evacuator tubing
and any connected attachment.
Figure 40 – To Remove Evacuator Components
8.4.1.3 To Remove Fluid Bag(s)
1. From the control screen menu, touch and hold the
IV pole DOWN button to lower the fluid bag(s). To
automatically lower the powered IV pole, push the
power switch to the OFF position (Figure 41).
2. Remove all fluid bag(s) from the IV pole hooks.
Figure 41 – To Remove Fluid Bag(s)
8.4.2 To Shut Down and Relocate the Rover
8.4.2.1 To Remove Power From the Rover
1. Push the power switch to the OFF position (Figure 42).
2. Disconnect the power cord of the rover from facility
electrical power. Wrap the power cord around the cord
bracket.
Figure 42 – To Remove Power