Suction ON
3760 mL
4 L suction stopped
4L Full
4L canister is full. 4L
suction has stopped.
Empty 4L canister to
restore suction.
Close
Empty 4L Canister
18.5
Figure 32 – Full Canister Condition
WARNING: SUCTION LOSS
– Suction is lost in both
canisters for at least 30 seconds during the EMPTY
4L CANISTER operation. Make sure the loss of
suction will not compromise patient safety.
If a 4-liter full canister condition occurs, perform one of
the following three options:
1. If the 20-liter canister can receive the contents of
the 4-liter canister, touch the EMPTY 4L CANISTER
button. An EMPTY 4L CANISTER dialog will appear
(Figure 33). Touch the CONFIRM button if the loss of
suction will not compromise patient safety.
Suction ON
Cancel
Confirm
Empty 4L
Canister
Empty 4L canister into
20L canister?
All suction will stop for
30 SECONDS while 4 L
canister empties.
Number of empties
remaining: 2
Figure 33 – Empty 4-Liter Canister Dialog
NOTE:
Status of the emptying and prefilling of the 4-liter
canister will be shown on the control screen (Figure 34
and Figure 35).
4450 mL
Emptying 4L
3450 mL
4 L suction stopped
20 L suction paused
Figure 34 – Emptying 4-Liter Canister Screen
7900 mL
Prefilling 4L
0 mL
4 L suction paused
20 L suction paused
Figure 35 – Prefilling 4-Liter Canister Screen