Maintenance Procedures – Litter
20
DISPLAYING INDIVIDUAL LOAD CELL OUTPUTS
A defective load cell can be detected by entering diagnostics and displaying individual load cell outputs.
1. To enter the diagnostic mode, unplug the bed power cord from the wall socket. Press and hold the LBS/KG
button. While holding the button, plug the bed power cord into the wall socket. Release the LBS/KG button
and the scale monitor will display “CALIBRATE SCALE”.
2. Press the arrow up button until the scale monitor displays “CHECK CORNERS”.
3. Press the ENTER button. The scale monitor should display “PICK CORNER NOW”.
4. Press and release the button that corresponds with the load cell to be checked (see below). The scale
monitor will display X/X=NNN.N. ”X/X” represents the initials of the selected corner, i.e. H/R will be dis-
played for the patient’s head end, right side. ”NNN.N” represents the load cell output.
5. Repeat step four for each corner. Head end weight readings will normally be lower than foot end weights.
Weight readings should be constant. A drifting 000.0 or 999.9 weight, or a reading that does not change
when weight is applied to that corner of the bed indicates a problem with the selected load cell assembly
or load cell cable.
6. When all the load cell outputs have been checked, press and release SCALE ON/OFF. To exit diagnostics,
unplug the bed’s power cord from the wall socket.
LOAD CELL REPLACEMENT
Required Tools:
9/16” Socket Wrench
9/16” Open End Wrench
Saw Horse (or Equivalent)
Wire Cutters
#2 Phillips Screwdriver
Replacement Procedure:
1. Raise the head or knee section, depending which end of the litter needs service.
2. Unplug the load cell connector from the load cell cable. If service is being done at the head end of the
bed, the J–channel must be removed to access the load cell connector.
3. Using wire cutters, remove the wire ties holding the cable to the frame.
4. Using a 9/16” socket and a 9/16” open end wrench, remove the two bolts holding the load cell to the litter
cross tube and remove the load cell.
5. Using a saw horse, support the litter at the end where the load cell was removed. Reverse the above
procedure to install the new load cell.
NOTE
Scale calibration procedure must be performed after the load cell is replaced (see page 21).
Summary of Contents for 2025
Page 42: ...Base Assembly 41 Assembly part number 2025 401 10 reference only FOOT END ...
Page 43: ...Base Assembly 42 HEAD END HEAD END ...
Page 44: ...Base Assembly 43 FOOT END ...
Page 57: ...2025 10 17 Rack Pinion Assembly Standard Height 2025 10 18 Rack Pinion Assembly Low Height 56 ...
Page 66: ...Litter Assembly 65 HEAD END Assembly part number 2025 231 10 reference only ...
Page 67: ...Litter Assembly 66 HEAD END ...
Page 69: ...Litter Assembly 68 HEAD END ...
Page 70: ...Litter Assembly 69 HEAD END ...
Page 74: ...Actuator Box Assembly 73 BY Assembly part number 2025 332 10 reference only ...
Page 75: ...Actuator Box Assembly 74 ...
Page 76: ...Actuator Box Assembly 75 ...
Page 77: ...Actuator Box Assembly 76 ...
Page 79: ...Actuator Box Assembly 78 DMS Options ...
Page 107: ...Foot Board Assembly 106 ...
Page 137: ...2025 40 10 Optional Bed Extender Mattress 136 ...