Maintenance Procedures – Base
16
MANUAL RELEASE VALVE REPLACEMENT
Required Tools:
#2 Phillips Screwdriver
Small Standard Screwdriver
Bungee Cords (or Equivalent)
Sawhorses (or Equivalent)
3/32” Allen Wrench
7/8” Open End Wrench
Replacement Procedure:
1. Raise the litter up to full height and unplug the bed power cord from the wall socket.
2. Support both ends of the litter top with sawhorses or the equivalent.
3. Using a #2 Phillips screwdriver, remove the three Phillips screws holding the base hood to the base frame.
4. If the bed is equipped with the 110V outlet option, remove the slotted screw holding the outlet cover. If
the bed does not have the 110V outlet option, remove the two slotted screws holding the cover plate.
5. Use bungee cords or the equivalent to secure the hood up and out of the way.
6. Using a 3/32” Allen wrench, loosen the set screw holding the release paddle to the pedal rod. Slide the
paddle out of the way.
7. Using a 7/8” open end wrench, remove the valve. Remove the stop sleeve from the old valve. It will be
used with the new valve.
8. Reverse the above procedure to install the new valve and reinstall the base hood.
SOLENOID RELEASE VALVE REPLACEMENT
Required Tools:
#2 Phillips Screwdriver
Small Standard Screwdriver
1 1/8” Deep Well Socket Wrench
Bungee Cords (or Equivalent)
Sawhorses (or Equivalent)
3/8” Nut Driver
5/32” Allen Wrench
Large Vise Grip Pliers
Wire Cutters
Replacement Procedure:
1. Remove the coil and wire from the solenoid (see page 15).
2. Support both ends of the litter top with sawhorses or the equivalent.
3. Using a 1 1/8” deep well socket wrench, remove the valve.
4. Reverse the above procedure to install the replacement valve.
HYDRAULIC PUMP COVER REMOVAL
Required Tools:
#2 Phillips Screwdriver
Standard Screwdriver
7/16” Socket Wrench
Bungee Cords (or Equivalent)
Removal Procedure:
1. Using a #2 Phillips screwdriver, remove the three Phillips screws holding the base hood to the base frame.
2. If the bed is equipped with the 110V outlet option, remove the slotted screw holding the outlet cover. If
the bed does not have the 110V outlet option, remove the two slotted screws holding the cover plate.
3. Use bungee cords or the equivalent to secure the hood up and out of the way.
4. Using a 7/16” socket wrench, remove the four bolts holding the hydraulic cover to the frame and remove
the cover.
Summary of Contents for 2025
Page 42: ...Base Assembly 41 Assembly part number 2025 401 10 reference only FOOT END ...
Page 43: ...Base Assembly 42 HEAD END HEAD END ...
Page 44: ...Base Assembly 43 FOOT END ...
Page 57: ...2025 10 17 Rack Pinion Assembly Standard Height 2025 10 18 Rack Pinion Assembly Low Height 56 ...
Page 66: ...Litter Assembly 65 HEAD END Assembly part number 2025 231 10 reference only ...
Page 67: ...Litter Assembly 66 HEAD END ...
Page 69: ...Litter Assembly 68 HEAD END ...
Page 70: ...Litter Assembly 69 HEAD END ...
Page 74: ...Actuator Box Assembly 73 BY Assembly part number 2025 332 10 reference only ...
Page 75: ...Actuator Box Assembly 74 ...
Page 76: ...Actuator Box Assembly 75 ...
Page 77: ...Actuator Box Assembly 76 ...
Page 79: ...Actuator Box Assembly 78 DMS Options ...
Page 107: ...Foot Board Assembly 106 ...
Page 137: ...2025 40 10 Optional Bed Extender Mattress 136 ...