Maintenance Procedures – Base
15
COIL AND WIRE REPLACEMENT (HEAD END)
Required Tools:
#2 Phillips Screwdriver
Small Standard Screwdriver
Bungee Cords (or Equivalent)
3/8” Nut Driver
7/64” Allen Wrench
Replacement Procedure:
1. Raise the litter up to full height and unplug the bed power cord from the wall socket.
2. Using a #2 Phillips screwdriver, remove the three Phillips screws holding the base hood to the base frame.
3. If the bed is equipped with the 110V outlet option, remove the slotted screw holding the outlet cover. If
the bed does not have the 110V outlet option, remove the two slotted screws holding the cover plate.
4. Use bungee cords or the equivalent to secure the base hood up and out of the way.
5. Using a 3/8” nut driver, remove the nut holding the coil and wire to the solenoid valve and remove the coil
and wire from the valve.
6. Using a 7/64” Allen wrench, remove the two screws holding the cover on the power junction box and re-
move the cover.
7. Unplug and remove the wires from the junction box.
8. Reverse the above procedure to install a new coil and wire assembly and reinstall the base hood.
COIL AND WIRE REPLACEMENT (FOOT END)
Required Tools:
#2 Phillips Screwdriver
Small Standard Screwdriver
Bungee Cords (or Equivalent)
3/8” Nut Driver
5/32” Allen Wrench
Large Vise Grip Pliers
Wire Cutters
Replacement Procedure:
1. Raise the litter up to full height and unplug the bed power cord from the wall socket.
2. Using a #2 Phillips screwdriver, remove the three Phillips screws holding the base hood to the base frame.
3. If the bed is equipped with the 110V outlet option, remove the slotted screw holding the outlet cover. If
the bed does not have the 110V outlet option, remove the two slotted screws holding the cover plate.
4. Use bungee cords or the equivalent to secure the hood up and out of the way.
5. Depress the pump pedal and hold it in the down position. Using large vise grip pliers, clamp the pump
bar from the top so the pliers rest against the pump motor plate.
NOTE
Be sure the pliers are clamped tight before continuing with the next step.
6. Using a 5/32” Allen wrench, loosen the two set screws on the pump pedal pivot bar.
7. Push the pivot bar completely out of the pump pedal assembly and lower the pedal assembly to the floor.
8. Using a 3/8” nut driver, remove the nut holding the coil and wire assembly to the solenoid valve and remove
the coil and wire from the valve.
9. Using wire cutters, cut the wire ties holding the wires to the frame and unplug the wires.
10. Reverse the above procedure to install the new coil and wire assembly
CAUTION
Do not remove the vise grips until the pump pivot bar is securely back in place. Damage to the equipment
or personal injury could occur.
Summary of Contents for 2025
Page 42: ...Base Assembly 41 Assembly part number 2025 401 10 reference only FOOT END ...
Page 43: ...Base Assembly 42 HEAD END HEAD END ...
Page 44: ...Base Assembly 43 FOOT END ...
Page 57: ...2025 10 17 Rack Pinion Assembly Standard Height 2025 10 18 Rack Pinion Assembly Low Height 56 ...
Page 66: ...Litter Assembly 65 HEAD END Assembly part number 2025 231 10 reference only ...
Page 67: ...Litter Assembly 66 HEAD END ...
Page 69: ...Litter Assembly 68 HEAD END ...
Page 70: ...Litter Assembly 69 HEAD END ...
Page 74: ...Actuator Box Assembly 73 BY Assembly part number 2025 332 10 reference only ...
Page 75: ...Actuator Box Assembly 74 ...
Page 76: ...Actuator Box Assembly 75 ...
Page 77: ...Actuator Box Assembly 76 ...
Page 79: ...Actuator Box Assembly 78 DMS Options ...
Page 107: ...Foot Board Assembly 106 ...