Hint
This sign indicates that additional information and hints are available.
2.3
Safety messages in this manual
CAUTION
Struers equipment must only be used in connection with and as described in the
instruction manual supplied with the equipment.
ELECTRICAL HAZARD
The machine must be earthed (grounded).
Make sure that the actual electrical power supply voltage corresponds to the
voltage stated on the type plate of the machine.
Incorrect voltage can damage the electrical circuit.
ELECTRICAL HAZARD
For electrical installations with Residual Current Circuit Breakers
For AbraPol-30 a residual current circuit breaker Type B, 30 mA is required (EN
50178/5.2.11.1).
For electrical installations without Residual Current Circuit Breakers
The equipment must be protected by an insulation transformer (double-wound
transformer).
Contact a qualified electrician to verify the solution.
Always follow local regulations.
CAUTION
Prolonged exposure to loud noises may cause permanent damage to a person’s
hearing.
Use hearing protection if the exposure to noise exceeds the levels set by local
regulations.
CAUTION
Risk of hand to arm vibration during manual preparation.
Prolonged exposure to vibration may cause discomfort, joint damage or even
neurological damage.
ELECTRICAL HAZARD
The pump of the recirculation cooling unit must be earthed (grounded).
Make sure that the electrical power supply voltage corresponds to the voltage
stated on the type plate of the pump.
Incorrect voltage can damage the electrical circuit.
CAUTION
The recirculation tank is very heavy when it is full.
2 Safety
AbraPol-30
9