6.4.3
Changing a preparation method
You can change an existing preparation method.
If you wish to change a Struers Metalog Guide methods, you must save the method under another
name and in a different group in the User Methods database before you can make any changes.
1.
Select the preparation method you wish to change.
2.
Go through the different preparation steps and make the changes as needed.
3.
Press
F4
.
–
Save method
Save the method with the current name and current method group.
–
Save method as
Specify a new method group, if needed, and a new method name.
6.4.4
Locking a preparation method
You can lock a preparation method to prevent it from being changed or deleted.
1.
In the
Group
screen, select the method you wish to lock.
2.
Press
F1
:
Lock method
.
3.
You will be prompted to confirm that you wish to lock the method.
4.
The symbol in front of the method name shows that the method is locked.
If you make changes to this preparation method, you must save the method
under another name.
6.4.5
Copying a preparation step
You can copy a preparation step from one preparation method to another.
1.
Select a method that includes the preparation step you wish to copy.
2.
Select the preparation step you wish to copy.
3.
Press
F1
:
Copy step
.
4.
Select the preparation method where you wish to insert the preparation step.
5.
Select the preparation step that will follow the new preparation step.
6.
Press
F2
:
Insert step
.
The new preparation step is inserted before the highlighted step.
7.
Press
F4
:
Save method
.
If the preparation method is locked, select
Save method as
6.4.6
Deleting a preparation step
You can delete a preparation step from a preparation method.
34
AbraPol-30
6 Operating the device