6.5.3
Adjusting the specimen holder position
To obtain the best preparation results, you must adjust the position of the specimen holder in
relation to the preparation disc.
1.
Release the handles located on each side of
the specimen holder motor.
The handles are spring-loaded to increase
the range of movement.
2.
Pull the handle outwards to disengage the
torque moment on the screw.
3.
Move the specimen holder motor sideways
manually until you have found the correct
position. You can read the eccentricity on the
scale on the specimen holder motor.
4.
Turn the handle clockwise (or counter-
clockwise) to the desired position, then
release the handle to re-engage the torque
moment on the screw.
5.
Tighten the screws again.
A
Handle
B
Screw
Further adjustments
For more precise adjustments, use a special function in the
Configuration
menu.
6.
From the
Main menu
select
Configuration
>
Specimen holder
positioning
.
7.
Press
F1
to move the specimen holder down. The specimen holder moves
down at the lowest possible force without rotation, and you can now position
the holder as precisely as required.
8.
Press
F1
again to move the specimen holder up.
6.6
The preparation process
6.6.1
The splash guard
Manual preparation
•
The splash guard for manual preparation is supplied with the
machine.
6 Operating the device
AbraPol-30
37