Insert the filter gasket in the coupling nut with the flat side against the pressure hose.
–
Tighten the pressure nut.
2.
Connect the other end of the pressure hose to the main water outlet
–
Mount the reduction ring with gasket on the cold water outlet if necessary.
–
Insert the gasket.
–
Tighten the coupling nut.
Adjusting the water flow
•
If needed, use the water valve to adjust the water flow
during grinding/polishing.
5.9
Connecting to the waste water outlet
1.
The machine is delivered with a water outlet hose.
2.
Lead the water outlet hose out of the machine through one of the openings in the sides of the
machine.
3.
Make sure that the hose slopes downward towards the waste water drain throughout its
entire length. If needed, shorten the hose.
5.10
Connecting the recirculation unit (option)
To ensure optimal cooling, mount a recirculation unit on the machine.
Note
Before you connect the recirculation unit to the machine, you must prepare it for
use. See the Instruction Manual for this unit.
ELECTRICAL HAZARD
Make sure that the electrical power supply voltage corresponds to the voltage
stated on the type plate of the pump.
Incorrect voltage can damage the electrical circuit.
5 Installation
AbraPol-30