Near zero is less than .04% of full scale capacity.
12
SPo oFF
Setpoint Output mode
Selects the control mode for the setpoint/comparator
outputs.
Select:
oFF - To disable setpoint or comparator operation.
The outputs will be in there de-activated
state as defined by the "SPA" parameter.
CutoF - To select setpoint/cutoff operation. The
setpoint output will actuate(with annunciator
indication) when the keyboard entered
setpoint value is exceeded.
COMP - To select comparator operation. This mode
is used for check weighing or any operation
requiring Over/Under/Go verification.
Outputs are controlled using the SP 1 and
SP 2 values such that: UNDER(SP 1) < GO <
OVER(SP 2). Upper and lower limits can be
set in the weigh or count mode. Front panel
annunciators indicate state of comparator
outputs.
13
SPA oFF
Setpoint Activation State
Sets the normal setpoint/comparator output state to
normally off(relay open) or normally on(relay
closed). Select:
oFF - If output is normally oFF(open)
on - If output is normally on(closed)
14
SPD qty
Setpoint Data Type
Sets setpoint/comparator actuation for quantity, net
weight, or gross weight. Select:
qty - Output will actuate on quantity data.
nEt - Output will actuate on net weight data.
grS - Output will actuate on gross weight data.
15
SPStbL F
Setpoint Activation on Stable Weight
Selects if setpoint/comparator outputs will activate
only on stable weight(no motion). Select:
F(alse) - If stable weight is not required.
t(rue) - If stable weight is required.
16
id-APW F
ID Memory APW configuration
Selects the information stored when ID memory is
used. Select:
F - No APW will be recalled with the stored ID
memory.
t - If the APW will be recalled along with total
quantity accumulation when the ID memory is used.
Sto
ID/Accumulation operation). Use this mode if APW