Page 31 of 54
tArE 1
tArE 1
In mode tArE1, pressing on the
key
,
stores the current displayed value unless it is
over scaled, the TARE LED will light up and
from this moment the value displayed is the
net value, i.e., the measured value minus the
value stored in the tare. Pressing on the
key for 3 s, the instrument will set the tare
value to zero, and the TARE LED will not
illuminate, indicating the GROSS value.
= Set tare (zero the display)
> 3 s = Clear tare and return the display to
normal operation
tArE 2
tArE 2
In this mode, a short press on the
key
has no effect. The tare value is introduced
manually. The edit menu can be accessed from
the “RUN” mode, by pressing on the
key
for 3 s. Following the diagram below.
= No effect
> 3 s = Manually enter a value for tare
(entering a zero will clear the tare function).
The display will show the measured value
minus the entered tare value.
tArE 3
tArE 3
In this mode, an unknown tare value can be
entered to give the required net value on the
display. This can be accessed from “RUN”, after
pressing on the
key for 3 s following the
diagram below. The LED TARE will become
active once the tare is applied by pressing the
key.