Operating Manual
13
1300-87076-1A Rev. 1
Table 7. Function icons.
Icon
Function
Description
Blank
To establish Blank data
Auto Run: On
The instruction of Auto Run function is on
Auto Run: Off
The instruction of Auto Run function is off
Measure
To do sample measurement or to do standard measurement
Delete
To delete sample measurement data
Note
: Standard measurement data cannot be deleted, but can be
overwritten
Save Result
To save the report
Back
Return to the last page
3.6.2 Basic Measurement Operation
1. Choose the correct
Method
type according to experiment.
2. Ensure the surface of the sample window and the cover window are both clean.
3. Adjust the pathlength selector to appropriate position/light path.
4. Mix the sample gently before adding it to the sample window.
5. Add appropriate solution of at least 1 µL and click
Blank
to establish blank data.
6. Wipe away the blank solution. Add the sample of at least 1 µL and click
Measure
.
7. Clean the sample window and cover window with a lint-free wipe between changing samples and after the
experiment is done. Use deionized water, ethanol or isopropanol if needed.
Note
: If the
Auto Run
function is on, the
Measure
will run automatically after closing the detection arm.